Afinion HbA1c Control 2 Levels

Item Number: 1463578

P.O.A. inc GST
Box
Quantity In Cart:  

NON RETURNABLE Please refer to our return policy for more information

Quality control testing using the Afinion HbA1c Control should be done to confirm that your Afinion Analyzer System is working properly and provides reliable results. Only when controls are used routinely and the values obtained are within acceptable ranges can accurate results for patient samples be assured

Frequency of control testing:

  • with each new shipment of Afinion HbA1c or Afinion HbA1c Dx Test Kits.
  • with each new lot of Afinion HbA1c or Afinion HbA1c Dx Test Kits.
  • at least every 30 days.
  • when training new operators in correct use of the Afinion HbA1c Test Kit and Afinion Analyzer.
  • anytime an unexpected test result is obtained. Consult the package insert for the Afinion HbA1c test in use, section “Interpretation of results”.

If local, state and/or federal regulations require more frequent testing of control materials, then quality control should be performed in compliance with these regulations.